Chronopotentiometry (CP)

Last Updated: 3/25/20 by Tim Paschkewitz

ARTICLE TAGS
  • AfterMath CP,
  • CC,
  • charge discharge,
  • chronopotentiometry,
  • constant current,
  • CP,
  • voltage vs. time
Article Contents/Section Navigation
  1. Technique Overview
  2. Fundamental Equations
  3. Experimental Setup in AfterMath
    1. Basic Tab
    2. Advanced Tab
    3. Ranges, Filters, and Post Experiment Conditions Tab
  4. Sample Data
  5. Applications from Literature
  6. References

1Technique Overview

Chronopotentiometry (CP) is a galvanostatic method also known as constant current (CC) in the battery field.  It is related to galvanostatic bulk electrolysis and has several variants, including Rotating Electrode (CP-RDE) and Rotating Cylinder (CP-RCE).

Chronopotentiometry (CP) is a galvanostatic method in which the current at the working electrode is held at a constant level for a given period of time. The working electrode potential and current are recorded as a function of time.  Researchers employ this method to study chemical reaction mechanisms and kinetics.  It is also frequently used to study batteries and electrodeposition.  CP is typically performed in an unstirred electrochemical cell; although, some researchers employ rotating electrodes in conjunction with CP.  Rotating Disk Chronopotentiometry (CP-RDE)
 
Typically, current is held constant between working and counter electrodes while potential is measured at the working electrode, relative to the reference electrode.  Redox-active species diffuse to the working electrode surface to balance the applied current, until the diffusion-limited concentration of redox species reaches zero at the electrode surface, at which time potential changes to the redox potential of the next species, if present, in solution (which could be solvent).  CP is also commonly used during battery charge and discharge experiments.
Back To Top

2Fundamental Equations

 
Consider the general reaction
 
\displaystyle O+ne^- \rightleftharpoons R
 
with formal potential E^{o'}.  As a constant current is applied to the working electrode, O is reduced to R at the working electrode surface.  As a result, the working electrode potential moves to values characteristic of the redox couple in a time-based Nernstian fashion.  As the concentration of O drops to zero at the electrode surface, the potential starts to rapidly increase to more negative values. The resulting E-t curve is much like a potentiometric titration with a transition time, \tau (analagous to titration equivalence point). The potential at one half \tau is E^{0'}.  The transition time is related to the concentration and diffusion coefficient of O through the expression
 
\displaystyle  \tau^{3/2} = {\frac{2C_0^*nFAD_0^{1/2}}{3\beta}}
 
where C_0^* is the concentration of O, n is the number of electrons, F is 96,485 C/F (Faraday's Constant), A is the working electrode area, D is the diffusion coefficient, and \beta is the sweep rate.
 
For rapid electrode kinetics, the time-based Nernstian equation is
 
\displaystyle E=E_{{\tau}/4}+\frac{RT}{nF}\; ln\left({\frac{\tau^{1/2}-t^{1/2}}{t^{1/2}}}\right)
 
where E_{\tau/4} is equal to
 
\displaystyle E_{\tau/4}=E^{0'}-\frac{RT}{2nF} ln \frac{D_O}{D_R}
 
A plot of E vs. \ln{\left({\frac{\tau^{1/2}-t^{1/2}}{t^{1/2}}}\right) } should give a straight line with slope of RT/nF for a reversible E-t curve.
 
Additional and more complete background of this theory is provided in the Theory section of the Knowledgebase.  Basic Methods
Back To Top

3Experimental Setup in AfterMath

To perform a chronopotentiometry experiment in AfterMath, choose Chronopotentiometry (CP) from the Experiments menu (see Figure 1).
 
Chronopotentiometry (CP) Experiment Menu Selection in AfterMath

Figure 1. Chronopotentiometry (CP) Experiment Menu Selection in AfterMath

 
Doing so creates an entry within the archive, called CP Parameters. In the right pane of the AfterMath application, several tabs will be shown (see Figure 2).
 
Chronopotentiometry (CP) Experiment in AfterMath

Figure 2. Chronopotentiometry (CP) Experiment in AfterMath

 
As with most Aftermath methods, the experiment sequence is
 
Induction Period → Electrolysis Period → Relaxation Period → Post-Experiment Idle Conditions
 
Unlike most experiments, the Induction and Relaxation Periods are on the Basic Tab.  The parameters for a CP experiment are fairly simple compared to other methods in AfterMath.
Back To Top

3.1Basic Tab

TIP:  Click the AutoFill button ("I Feel Lucky" prior to May 2019) on the top bar in AfterMath to automatically fill all required parameters with reasonable starting values.  While the values provided may not be appropriate for your specific system, they are reasonable parameters with which to start your experiment, especially if you are new to the method.

 
The basic tab contains fields for the fundamental parameters necessary to perform a CP experiment.  AfterMath shades fields with
yellow when a required entry is blank and shades fields pink when the entry is invalid (see Figure 3).
 
Chronopotentiometry (CP) Experiment Basic Tab in AfterMath

Figure 3. Chronopotentiometry (CP) Experiment Basic Tab in AfterMath

 
During the induction period,  Induction Period a set of initial conditions are applied to the electrochemical cell and the cell equilibrates at these conditions.  Data are not collected during the induction period, nor are they shown on the plot during this period.
 
After the induction period, the current applied to the working electrode is stepped to the specified value for the duration of the experiment, which is called the electrolysis period.  The galvanostatic circuit of the instrument maintains constant applied current while simultaneously measuring the potential at the working electrode relative to the reference electrode.  During the electrolysis period, potential and current at the working electrode are recorded at regular intervals as specified on the Basic tab.
 
The experiment concludes with a relaxation period.  Relaxation Period  During the relaxation period, a set of final conditions (specified on the Advanced tab) are applied to the electrochemical cell and the cell equilibrates at these conditions.  Data are not collected during the induction period, nor are they shown on the plot during this period.
 
A plot of the typical experiment sequence, containing labels of the fields on the Basic tab, helps to illustrate the sequence of events in a CP experiment (see Table 1 and Figure 4).

Field labels in AfterMath will eventually include symbols, such as those used in the waveform plot below.  Consult the cross-reference table (see Table 1) to match field names with symbols.

Group Name Field Name Symbol
Induction Period Current I_{IND}
Induction Period Duration T_{IND}
Electrolysis Period Current I_{E}
Electrolysis Period Duration T_{E}
Relaxation Period Current I_{RLX}
Duration Duration T_{RLX}
Sampling Control Number of Intervals N_{INT}
Table 1.  Basic Tab Group Names, Field Names, and Symbols.
 
ChronopotChronopotentiometry (CP) Basic Tab Field Diagramentiometry (CP) Basic Tab Field Diagram

Figure 4. Chronopotentiometry (CP) Basic Tab Field Diagram

 
At the end of the relaxation period, the post-experiment idle conditions Understanding Post Experiment Conditions Tab are applied to the cell, and the instrument returns to the idle state.  The default plot generated from the data is measured potential vs. time, called a chronoamperogram.
Back To Top

3.2Advanced Tab

The CP Advanced tab contains two groups for iR Compensation and for Experiment End Trigger (see Figure 5).
 
Chronpotentiometry (CP) Experiment Advanced Tab in AfterMath

Figure 5. Chronpotentiometry (CP) Experiment Advanced Tab in AfterMath

 
Detailed description of the iR Compensation Mode is provided elsewhere on the knowledgebase.  iR Compensation This mode is used to correct for uncompensated resistance in the electrochemical cell.
 
In a CP experiment, the researcher might want to have AfterMath monitor the response (input) and then stop the experiment a specific current, potential, and/or charge.  This value is called a trigger and is set in the Experiment End Trigger group on the Advanced tab (see Figure 5).  For CP, the signals for the trigger can be potential or current.  A common example of using this trigger is in the charging of a battery.  A researcher would charge the battery by performing a CP experiment (also called a CC or constant current)  with the trigger set to its upper potential limit (e.g., 100% SOC).  Once this potential limit is reached, the CP experiment terminates and retains the data collected up to the trigger endpoint.  More information about end triggers is available on the Knowledgebase. End Triggers
Back To Top

3.3Ranges, Filters, and Post Experiment Conditions Tab

The Ranges tab contains the Electrode Range group, which contains the initial current and potential range fields.  Nicotinic acetylcholine receptor (nAChR) mediated dopamine release in larval Drosophila melanogaster These fields are often duplicated, and linked to, the same fields on the Basic tab.  Additionally, a field for Autorange control for both potential and current appear on this tab.   More information on autorange is available elsewhere on the knowledgebase.  Application of Measurement Models to Impedance Spectroscopy
 
The Filters tab provides access to potentiostat hardware filters, including stability, excitation, current response, and potential response filters.  Pine Research recommends that users contact us  Contact for help in making changes to hardware filters.  Advanced users may have an easier time changing the automatic settings on this tab.  Filter settings fields are shown for WK1 (working electrode #1) as well as for WK2 (working electrode #2) regardless of the potentiostat connected to AfterMath.  More information on filters is available elsewhere on the knowledgebase.  Drosophila as a Model System for Neurotransmitter Measurements
 
By default, the potentiostat disconnects from the electrochemical cell at the end of an experiment.  There are other options available for what these post-experiment conditions can be and are controlled by setting options on the Post Experiment Conditions tab.  Complete details on the fields and settings on this Post Experiment Conditions tab are provided elsewhere on the knowledgebase.  Understanding Post Experiment Conditions Tab
 
Back To Top

4Sample Data

The typical chronopotentiogram for a 2.5 \; mM solution of Ferrocene in 0.1 \; M \; Bu_4NClO_4(MeCN) shows a plot similar to a potentiometric titration (see Figure 6). Initially, the potential is slightly rising until nearly all of the ferrocene at the electrode is consumed, at which point the potential rises rapidly through an inflection point, followed by a trailing off of the potential after the inflection point.  AfterMath has many ways to view and analyze data as well as to customize plots.  These additional aspects of AfterMath are detailed elsewhere in the knowledgebase. 
 
Chronopotentiogramm of Ferrocene

Figure 6. Chronopotentiogramm of 2.5 mM Ferrocene

 
The addition of a Baseline Tool  Ultra-Capacitive Carbon Neural Probe Allows Simultaneous Long-Term Electrical Stimulations and High-Resolution Neurotransmitter Detection aids in finding the inflection point, the point at which the slope of the plot is greatest (see Figure 7).  Once the tool has been added to the trace on the plot, right-click the tool to view its properties.  There, change the baseline mode to Tangent to easily find the inflection point.  More on the use of Aftermath tools are found within the knowledgebase.   The potential at {1/2}\tau \approx 12.3s of the inflection point is approximately the E^0 for the species of interest ( 0.553 \; V ). Two additional Crosshair  Real-Time Detection of Melatonin Using Fast-Scan Cyclic Voltammetry tools have been added to the plot to show the time at the inflection point and the time of {1/2}\tau of the inflection point.
 
Chronopotentiogram of Ferrocene with Inflection Point Labeled

Figure 7.  Chronopotentiogram of Ferrocene with Inflection Point Labeled

Back To Top

5Applications from Literature

Here, CP was used to measure diffusion coefficients.   Le et al. generated concentration polarization at an anion exchange membrane under constant current (CP).  Le, X. T.; Viel, P.; Tran, D. P.; Grisotto, F.; Palacin, S.  Surface Homogeneity of Anion Exchange Membranes: A Chronopotentiometric Study in the Overlimiting Current Range.  The Journal of Physical Chemistry B, 2009, 113(17), 5829-5836.  The potential drop across the membrane was monitored as a function of time.  By determining the transition time (point of inflection) calculate diffusion coefficients were calculated for several chloride salts in 0.1\; M solutions. An exquisite example, considering the authors were able to determine diffusion coefficients of non-redox active species.
 
In another example,  Nagaraju and Lakshminarayanan used CP to grow mesoporous Au films for use in the electro-oxidation of alcohols in alkaline media.  Nagaraju, D. H.; Lakshminarayanan, V.  Electrochemically Grown Mesoporous Gold Film as High Surface Area Material for Electro-Oxidation of Alcohol in Alkaline Medium.  The Journal of Physical Chemistry C, 2009, 113(33), 14922-14926. The researchers showed that varying the current density and deposition yielded different surface areas.  These films were then used to electro-oxidize methanol or ethanol indicating that they could be used in direct alcohol alkaline fuel cells.
Back To Top

6References

Back To Top
Questions? Find Answers.

Our knowledgebase is the central repository for written content, including help topics, theory, application notes, specifications, and software information.

Select a category below or view the Knowledgebase home page
Software

Detailed information about our Software, which includes AfterMath and retired PineChem.

Applications

Application notes discuss practical aspects of conducting specific experiments.

Theory

Fundamental electrochemical theory presented in a brief and targeted manner.

Product Specifications

Review complete product specifications and compare products within a category here.