Reference Electrode, Silver/Silver Chloride, Single Junction

Part Number
RREF0021

The standard-sized, single junction, silver/silver chloride (Ag/AgCl) reference electrode is filled with 4M KCl with AgCl solution. The refrerence electrode body is 9.5 mm OD, therefore ideally used with Standard volume electrochemical cells. The reference electrode is designed to be pass through a 14/20 ground glass joint and held in place by the included PTFE adapter, which can be adjusted to control the immersion depth of the reference electrode into the cell. Pine Research recommends proper storage and use of reference electrodes. A variety of useful tools for storage and care are shown below.

Login to View Prices

Customers must be logged into their account to view prices. Not all regions provide pricing online. If you do not see prices, you can obtain them from the designated sales channel in your region.

The standard-sized, single junction, silver/silver chloride (Ag/AgCl) reference electrode is filled with 4M KCl with AgCl solution. The refrerence electrode body is 9.5 mm OD, therefore ideally used with Standard volume electrochemical cells. The reference electrode is designed to be pass through a 14/20 ground glass joint and held in place by the included PTFE adapter, which can be adjusted to control the immersion depth of the reference electrode into the cell. Pine Research recommends proper storage and use of reference electrodes. A variety of useful tools for storage and care are shown below.

Important
Specifications
References
Documents
SDS
Compatibility Notice

Avoid using this reference electrode with the following: NH3 buffers, sulfides, perchlorates, non-aqueous/organic solvents, and alkaline solutions stronger than 0.1M OH.

Thermal Stability Range: 10°C - 80°C

Electrode damage may occur at temperatures beyond recommended range. This range is provided for electrodes with PEEK shrouds.

Reference Electrodes
Reference chemistry
Ag/AgCl
Standard potential
+199 mV vs NHE
Reference size type
Reference electrode body diameter
Reference electrode length
19 cm
Reference frit material
External fill solution
4 M KCl with AgCl
Refillable
Typical variance
±3 - 5 mV
Typical input impedance
< 10 kΩ
Reference electrode connection style
1.6 mm OD x 8.5 mm long brass pin
Document #
Title
Type
Download
DRP10031
Reference Electrode Storage and Master Reference Electrodes
DRK10053
Overview of Reference Electrode Operation and Alternative Reference Electrodes
For this product, any applicable SDS will be listed here by category. You can view or download the SDS by clicking the appropriate button.
Reference Electrode Filling Solutions
4 M Potassium Chloride with Silver Chloride (KCl with AgCl)