The WaveNeuro Fast Scan Cyclic Voltammetry (FSCV) Potentiostat System is a unique electrochemical instrument you will not find from other traditional electrochemical suppliers.
Grounding
Signal Ground (DC Common)
Chassis Terminal
The signal ground is isolated from the power plug and floats with respect to the instrument chassis and earth ground.
The chassis terminal is a banana binding post (back panel) which may optionally be used to connect the chassis to earth ground to improve noise screening (shielding).
Measured Current
Current Range
Resolution
Aux Input Resolution
Filters
Defined by electrochemical headstage – (±200 nA/V headstage amplifier included)
16-bit resolution (provided by the interface board)
16-bit resolution (provided by the interface board)
Selectable: low-pass up to 14 kHz, low-pass up to 40 kHz, bandpass (100 Hz to 14 kHz) or bypass (no filtering)
Applied Potential (Potential Ramp WaveForm)
Ranges
Resolution
DAC Output
Filters
CV Scan Rate (max)
±3.3 V (maximum range)
100 μV (minimum resolution defined by the interface board)
16-bit resolution (provided by the interface board)
Selectable; lowpass up to 2 kHz, 5 kHz or 10 kHz, and bypass (no filtering)
5,000 V/s
Measured Potential In (front panel labeled as NON-CV IN)
Ranges
Resolution (at each range)
ADC Input
±10.0 V (maximum range defined by the interface board)
313 μV (minimum resolution defined by the interface board)
16-bit resolution (provided by the interface board)
Applied Potential Out (front panel labeled as ABS OUT)
Ranges
Resolution (at each range)
ADC Input
±10.0 V (maximum range defined by the interface board)
313 μV (minimum resolution defined by the interface board)
16-bit resolution (provided by the interface board)
Data Acquisition
Point Interval
Synchronization
Raw Point Total
500 ns (minimum)
simultaneous sampling of all analog input signals
>10 million per experiment
Accessories
Dummy Cell
Headstage Cable
Electrode Connections
Power Supply
Headstage Amplifier
external dummy cell (included)
DB-25 to Pine Research electrochemical headstage
white = reference; yellow = working
C14-type socket (includes plug-specific 10A cord)
200 nA/V, working electrode driven (other gains available by special order)
Output Connections (Top Panel)
All output connections to the WaveNeuro are via female BNC-type connectors.
STIM +
STIM –
CVF
SPARE
FLOW
E-PHYS
STIM OUT
Digital, TTL compatible
Digital, TTL compatible
CV frequency; Digital, TTL compatible
Digital, TTL compatible
Flow cell trigger; Digital, TTL compatible
Timing Trigger; Digital, TTL compatible
Stimulus waveform; Analog, max ±10.0 V
Input Connections (Top Panel)
PASSTHRU
BEHAVIORAL
External stimulus to headstage input; Analog, BNC female
Female DB-25 connector for recording up to 15 TTL signals
Interface Board Connections
Connector 0
Connector 1
Compatible DAQ Interface
DAQ Interface Cable
Control Software
68 –pin VHDCI male connector
68 –pin VHDCI male connector
National Instruments PCIe-6363 (sold separately)
National Instruments SHC8-68 EPM x 2
Industry standard LabVIEW based software available to researchers (not provided)
General Specifications
Power Required
Power Cord
LED Indicators
Instrument Dimensions
Instrument Weight
Temperature Range
Humidity Range
Power Supply
12.0 VDC (±5%)
various international cables available (10A, C13 Connector)
power, input and output filter selection, trigger activity
58× 242 × 242 mm (2.25 × 9.5 × 9.5 in)
2.0 kg (1.8 lb)
10˚C to 40˚C
80% RH maximum, non-condensing
input: 100 to 240 VAC,2.3 A,50 to 60 Hz
output: 12.0 VDC,2.5 A