WaveNeuro Four Channel Headstage Kit

Expand your FSCV capabilities with our four-channel headstage kit.  Pine Research released the WaveNeuro Four in 2018 and until now, users would have to use four individual headstage cable kits, each having their own headstage amplifier and electrode connections.  The new Four-Channel Headstage Kit solves this issue by providing a single headstage that accommodates up to four channels simultaneously to a common reference electrode.  The features of this new product include the following:

  • Up to four working electrode channels, each color coded
  • All four channels have the same gain, 200 nA/V (5 MΩ)
  • Simple connections – only one cable from WaveNeuro Four to terminal electrodes
  • Novel piecewise design – identify issues and replace individual components as needed

Product Information

Important Product Warnings and Notices

Working Driven Headstage
Pine Research currently offers working driven headstage amplifiers. In a working driven system, the reference electrode is grounded. The FSCV potential waveform (ramp) is connected to the non-inverting input of the operational amplifier, while the working electrode is connected to the inverting input. In this arrangement, the voltage at the microelectrode will follow the ramp applied to the inverting input. This means that even when no electrodes are connected (connector just in air) to the headstage amplifier cable, which is connected to the WaveNeuro, HDCV will show a current response that follows the applied waveform, which is the expected behavior for a working driven headstage.
FSCV Instrumentation Journal Article
R.M. Wightman et. al. have well-described FSCV instrumentation, including working driven headstages. Pine Research has based the design of the WaveNeuro FSCV systems and accessories in a similar manner as that which they decribe.

Additional Product Details

With the WaveNeuro Four Channel Headstage Kit, users can more conveniently connect up to four working electrodes to a single headstage.  To make this possible, the Four Channel Headstage Kit includes an adapter board, to which a special interface cable and four-channel headstage cable connects.  Terminating this assembly are the four-channel headstage-to-microelectrode coupler wires, which includes sockets for four working electrodes (Yellow, Red, Blue, and Green) and one reference electrode (White). The Four Channel Headstage Kit finds application in several areas, yet to be fully realized by the FSCV research community:

  • Easy throughput. Deliver the same waveform to four channels simultaneously, and measure current at each channel independently.  In this case, you might install the electrodes into the same region of the brain, same tissue, or same flow cell and gain n=4 with a single experiment.
  • Multiple analytes.  Imagine measuring multiple analytes in your system simultaneously.  Look at dopamine, serotonin, adenosine, and other molecules simultaneously – simply apply the appropriate waveform to each channel independently.

The FSCV community has yet to release seminal papers on the topic of multichannel FSCV.  Pine Research hopes to promote the growth of FSCV into this area by providing a well-designed commercial solution for those needing multichannel FSCV.

Pine Research currently offers working driven headstage amplifiers.  In a working driven system, the reference electrode is grounded.  The FSCV potential waveform (ramp) is connected to the non-inverting input of the operational amplifier, while the working electrode is connected to the inverting input.  In this arrangement, the voltage at the microelectrode will follow the ramp applied to the inverting input.1

In this two-electrode configuration, current arising from electron-transfer reactions, such as the oxidation of dopamine, passes between reference and working electrodes.  The measured current passes through the headstage amplifer, where it is converted to voltage, and sums with the the ramp voltage at the inverting input. Mathematically,

\displaystyle{V_O=-(i_{in}\times R_F)+V_R}

where V_O is the output voltage, i_{in} is input current,
R_F is feedback resistor (gain), and V_R is the CV ramp voltage.  By rearrangement, the signal voltage (proportional to the current across the 5 MΩ feedback resistor in the 200 nA/V headstage) is then

\displaystyle{(V_R-V_O)=i_{in}\times R_F=V_{signal}}

HDCV software, which supports the WaveNeuro FSCV Potentiostat system, performs software subtraction of the ramp according to this relationship, resulting in only the true differential current measurement.1

FSCV Instrumentation Journal Article
R.M. Wightman et. al. have well-described FSCV instrumentation, including working driven headstages. Pine Research has based the design of the WaveNeuro FSCV systems and accessories in a similar manner as that which they decribe.

Product Specifications

The specifications of this product, and any other product that may be included in a kit or bundle, or a part that is included with this product, are listed below.
WaveNeuro Four Channel Headstage Kit

Product Kits

Product kit
Yes
Kit includes
Cables, Amplifier, Adapter(s), Circuit Board Interface

Products Included in this Kit

Product kits contain more than one product. Should you want more details or wish to order a replacement component, visit the product page below.

Optional Product Add-Ons

Customers may wish to order optional add-on products and accessories compatible with this product. Should you want more details or wish to order an additional item, refer to the list below.

Request a Quote for this Cart

Below is the billing data we have on file. These details will be used on your quotation, so please update them here if you require any changes. Changing any details here updates them for your account.