WaveDriver 10 DC Potentiostat
WaveDriver 10 Potentiostat/Galvanostat System
Product Discontinued - No Replacement Available
This product has been discontinued and can no longer be purchased. The product remains on our website for reference and a listing of its specifications. There is no direct product replacement to offer.
Product Information
Product Specifications
The specifications of this product, and any other product that may be included in a kit or bundle, or a part
that is included with this product, are listed below.
WaveDriver 10 DC Potentiostat
Electrochemical Workstations
Communication
Interface
USB
Wireless capable
No
General
Power input
24.0 VDC (±5%), 4.0 A (low voltage DC device)
Power supply input
100 - 240 VAC, 1.4 - 0.7 A, 50 - 60 Hz
Power supply output
24 VDC, 5.0 A power supply (included) has a C14 type input connector
Power cord
Various international cables sold separately (C13 type)
LED indicators
Power, USB, and status
Instrument dimensions
160 × 324 × 255 mm (6.3 × 12.75 × 10.0 in)
Workstation shipping dimensions
254 × 356 × 457 mm (10 × 14 × 18 in)
Instrument weight
4.6 kg (10.2 lb)
Workstation shipping weight
7.7 kg (17 lb)
Temperature range
10°C - 40°C
Humidity range
80% RH maximum, non-condensing
Workstation modes
Potentiostat (POT), Galvanostat (GAL), Open-Circuit Potential (OCP), Zero-Resistance Ammeter (ZRA)
Auxiliary Connections
Connector C
9-pin DSUB connector includes digital signal ground, two digital output signals, and three digital input signals
Trigger input
BNC female, TTL compatible
Trigger output
BNC female, TTL compatible
Potential (E1) output
N/A
Current (I1) output
N/A
Potential (E2) output
N/A
Current (I2) output
N/A
Auxiliary analog input
N/A
Auxiliary analog output
BNC female, ±10 V bipolar output, 313 µV resolution, 0.2% accuracy (available when second working electrode not in use)
WK1 input
BNC female, ±10V differential input, 20 kΩ impedance, ±0.5% accuracy; allows external waveform to be summed directly to the working electrode excitation signal
WK2 input
BNC female, ±10V differential input, 20 kΩ impedance, ±0.5% accuracy; allows external waveform to be summed directly to the working electrode excitation signal
Accessories
Dummy cell
External dummy cell included
Cell cable
Combination D-SUB connector to multiple banana plugs via shielded coaxial cables (included)
Rotator Control Connections
Rotator connector A
7-pin mini circular DIN includes analog and digital signal grounds, digital rotator enable signal, auxiliary digital output signal, and analog rotation rate control signal
Rotator connector B
3-pin connector includes analog signal ground, digital rotator enable signal (+15 V max), and analog rotation rate control signal
Rate control signal
±10 V, ±2.5 V
Digital enable signal
open drain (TTL compatible)
Electrochemical Impedance Spectroscopy (EIS)
EIS capable
No
EIS frequency range
N/A
EIS frequency resolution
N/A
EIS frequency stability
N/A
Modes
N/A
Voltage excitation setpoint
N/A
Current excitation setpoint
N/A
Frequency sweeping
N/A
EIS accuracy
N/A
Data Acquisition (for DC Experiments)
Clock resolution
10 ns (minimum time base)
Point interval
80 µs (minimum)
Synchronization
Simultaneous current and potential input
Raw point total
<10 million per experiment
Applied Potential (Potentiostatic Mode)
Potential ranges (applied)
±10 V, ±15 V
Potential resolution at each range (applied)
469 µV per DAC bit, 313 µV per DAC bit, 78 µV per DAC bit
DC accuracy (potential, applied)
±0.2% of setting; ±0.05% of range
DAC output (potential)
16 bits
CV sweep rate (minimum)
10 µV/s
CV sweep rate (maximum)
125 V/s
Measured Potential
Potential ranges (measured)
±15 V, ±10 V
Potential resolution at each range (measured)
469 µV per ADC bit, 313 µV per ADC bit
DC accuracy (potential, measured)
±0.2% of setting; ±0.05% of range
ADC output
16 bits
Filters (for DC Experiments)
10 kHz, 1 kHz, 100 Hz, 30 Hz, 10 Hz
Electrometer (Reference Electrode Amplifier)
Input impedance
>10¹³ in parallel with <10 pF
Input current
<10 pA leakage/bias current at 25°C
CMRR
>100 dB, 0 - 1 kHz, >74 dB at 10 kHz
Electrometer bandwidth
>11 MHz (3 dB)
Power Amplifier (Counter Electrode Amplifier)
Output current
±1 A (maximum)
Short circuit current limit
undetermined
Compliance voltage
±16.5 V
Bandwidth
>200 kHz (on fastest speed setting)
Noise and ripple
undetermined
Slew rate/rise time
10 V/µs (on fastest speed setting)
Applied Current (Galvanostatic Mode)
Current ranges (applied)
±1 A, ±100 mA, ±10 mA, ±1 mA, ±100 µA, ±10 µA, ±1 µA, ±100 nA
Current resolution at each range (applied)
31.3 µA, 3.13 µA, 313 nA, 31.3 nA, 3.13 nA, 313 pA, 31.3 pA, 3.13 pA
DC accuracy (current, applied)
±0.2% of setting; ±0.05% of range
DAC output (current)
16 bits
Measured Current (Potentiostatic Mode)
Current ranges (measured)
±1 A, ±100 mA, ±10 mA, ±1 mA, ±100 µA, ±10 µA, ±1 µA, ±100 nA
Current resolution at each range (measured)
31.3 µA, 3.13 µA, 313 nA, 31.3 nA, 3.13 nA, 313 pA, 31.3 pA, 3.13 pA
Autoranging
Yes
Practical current range
100 pA to 1 A
DC accuracy (current, measured)
±0.2% of setting; ±0.05% of range
DC leakage current
<10 pA at 25°C
AC accuracy (measured)
N/A
AC leakage current
N/A
ADC input
16 bits
Filters (for DC Experiments)
10 kHz, 1 kHz, 100 Hz, 30 Hz, 10 Hz
Ground Connections
DC common (signal)
The DC Common is isolated from the USB port, the instrument chassis and earth ground. The DC Common is accessible via a banana binding post (black) on the back panel.
Chassis terminal
The metal case (chassis) terminal is a banana binding post (back panel) which may optionally be used to connect the chassis to earth ground or signal ground to improve noise screening (shielding).
Earth
No direct connection to earth ground is provided.
Electrode Connections
Reference electrode
Sense line with driven shield
Counter electrode
Drive line with grounded shield
Working electrode channels
1 Channel
Working electrode #1 (WK1)
Separate sense and drive lines, each with driven shield (current measurement via passive shunt)
Publications Citing this Product
Whenever possible, we add references to this list where we know this product was used. Do you have a
publication using this product but do not see your publication listed? Let us know and we will be happy
to add it!
- He, S.; Kapur, A.; Pakhomenko, E.; Rofshus, R.; Holmes, R.J. Engineering Spontaneous Orientation Polarization in Blends of Polar Organic Semiconductors. Adv. Opt. Mater. 2025, e01416.
- Igba, V.M.; Garcia-Lobato, M.A.; García-Pérez, U.M.; Oyervides-Muñoz, E.; Martínez-Mora, E.I. Activated carbon fibers derived from dryer lint via self-generated atmosphere for supercapacitor applications. Ionics 2024, 30, 5853-5860.
- Kayode, S.E.; Awobifa, O.S.; Garcia-Lobato, M.A.; Rosas, M.T.; Hoyos, M.; González, F.J. In Situ PANI–Graphite Nanochain-like Structures and Their Application as Supercapacitive Electrodes. Journal of Composites Science 2024, 8, 200.
- Garcia-Lobato, M.A.; Mtz-Enriquez, A.I.; Garcia, C.R.; Velazquez-Manzanares, M.; Avalos-Belmontes, F.; Ramos-Gonzalez, R.; Garcia-Cerda, L.A. Corrosion resistance and in vitro bioactivity of dense and porous titania coatings deposited on 316L SS by spraying method. Appl. Surf. Sci. 2019, 484, 975-980.
- Garcia-Lobato, M.A.; Garcia, C.R.; Mtz-Enriquez, A.I.; Lopez-Badillo, C.M.; Garcias-Morales, C.; Muzquiz-Ramos, E.M.; Cruz-Ortiz, B.R. Enhanced electrochromic performance of NiO-MWCNTs thin films deposited by electrostatic spray deposition. Mater. Res. Bull. 2019, 114, 95-100.
- Lydon, B.R.; Lee, C.C.; Tanifuji, K.; Sickerman, N.S.; Newcomb, M.P.; Hu, Y.; Ribbe, M.; Yang, J. Electrochemical characterization of isolated nitrogenase cofactors from Azotobacter vinelandii. Chembiochem 2019.
- Lydon, B.R.; Germann, A.; Yang, J.Y. Chemical modification of gold electrodes via non-covalent interactions. Inorg. Chem. Front. 2016, 3, 836-841.
- Wang, Y.; Zhao, N.; Fang, B.; Li, H.; Bi, X.T.; Wang, H. A highly efficient PtCo/C electrocatalyst for the oxygen reduction reaction. RSC Adv. 2016, 6, 34484-34491.
- Gunawardhana, D.Y.R.; Kaumal, M.N. Development of a portable paper-based microfluidic device for the detection of alcohol in biological fluids. Sri Lankan Journal of Biology 2016, 1, 38.
- Thammavongsy, Z.; Khosrowabadi Kotyk, J.F.; Tsay, C.; Yang, J.Y. Flexibility is Key: Synthesis of a Tripyridylamine (TPA) Congener with a Phosphorus Apical Donor and Coordination to Cobalt(II). Inorg. Chem. 2015, 54, 11505–11510.
- Tsay, C.; Livesay, B.N.; Ruelas, S.; Yang, J.Y. Solvation Effects on Transition Metal Hydricity. J. Am. Chem. Soc. 2015, 137, 14114–14121.
- Wang, Y.; Zhao, N.; Fang, B.; Li, H.; Bi, X.T.; Wang, H. Effect of different solvent ratio (ethylene glycol/water) on the preparation of Pt/C catalyst and its activity toward oxygen reduction reaction. RSC Adv. 2015, 5, 56570-56577.
- Tawfic, A.F.; Dickson, S.E.; Kim, Y.; Mekky, W. Enhanced Capacity and Stability for the Separation of Cesium in Electrically Switched Ion Exchange. Fusion Science and Technology 2015, 67, 608–611.
- Armutlulu, A.; Bottomley, L.A.; Bidstrup Allen, S.A.; Allen, M.G. Supercapacitor Electrodes Based on Three-Dimensional Copper Structures with Precisely Controlled Dimensions. ChemElectroChem 2015, 2, 236–245.
- Shaffer, D.W.; Johnson, S.I.; Rheingold, A.L.; Ziller, J.W.; Goddard, W.A.; Nielsen, R.J.; Yang, J.Y. Reactivity of a Series of Isostructural Cobalt Pincer Complexes with CO2 , CO, and H+. Inorg. Chem. 2014, 53, 13031–13041.
- Tsang, M.; Armutlulu, A.; Martinez, A.; Herrault, F.; Allen, S.A.B.; Allen, M.G. A MEMS-enabled biodegradable battery for powering transient implantable devices. 2014 IEEE 27th International Conference on Micro Electro Mechanical Systems (MEMS) 2014, 358–361.