This FAQ applies to versions 1.0.2447 and greater.[/call_to_action]
Attempting to perform an experiment from AfterMath on the Pine AFCBP1 bipotentiostat using the ISA-type National Instruments card (AT-MIO-16E-10) may result in the following error:
This problem can be caused by a lack of DMA resources. This can be confirmed using either AfterMath or the National Instruments Measurement & Automation Explorer as shown below:
The original cause of this failure occurs during the installation of the National Instruments interface board device drivers, and can also be seen in the Windows Device Manager:
The problem can be fixed by using the Windows “Device Manager” as shown below.
See also the National Instruments website article about this problem here
End User Certification - Checkout Form (/checkout)
All orders from Pine Research Instrumentation that are destined to be exported for use outside of the United States requires that the parties involved (end-user, intermediate consignee, and ultimate consignee) disclose the parties involved in the transaction. By signing this form, parties acknowledge having read and understood export restrictions and requirements.