1. Technique Overview
Normal Pulse Voltammetry (NPV) is a potentiostatic method that offers some advantages to common techniques like Cyclic Voltammetry (CV), in that the waveform is a series of pulses that are increasing in potential amplitude. The way in which the current is measured at each pulse aids in minimizing the measurement of background (charging) current. Related pulse-type methods include Differential Pulse Voltammetry (DPV) and Square Wave Voltammetry (SWV).
INFO:
Initial sweep direction can be positive-increasing or negative increasing. Not all possibilities have been provided as examples in this article. Users must tailor their parameters to suit their specific electrochemistry. Contact us with any questions or additional assistance.
2. Fundamental Equations
Consider the reaction
(6)
where
is reduced in a one-electron reaction to
with formal potential
. The application of a baseline potential should be sufficiently positive of
such that no faradaic current flows. After a period of time
, typically 100 - 5000 ms, the potential of the working electrode is stepped to a more negative value for a period of time. The total time from the application of the baseline potential through the application of the potential pulse is
.
The potential pulse is incrementally increased with each cycle. As the potential of the working electrode approaches
faradaic current flows due to the reduction of
to
. Upon the application of the baseline potential in the next cycle,
is oxidized back to
. When the potential of the working electrode gets sufficiently negative of
,
is reduced to
at a maximum rate and the current levels off to a plateau. The magnitude of this current plateau is given by
(7)
where the parameters are as described above. Notice that in the Sample Experiments section, there is a slight anodic current flowing at the beginning of the experiment. The magnitude of this current is given by the equation
(8)
where the parameters are as described above. The magnitude of this current is the difference between the NPV and RNPV currents.
Bard and Faulkner1 provide a summary and description of normal pulse voltammetry as do Kissinger and Heinemann2.
3. Experimental Setup in AfterMath
To perform a normal pulse voltammetry experiment in AfterMath, choose Normal Pulse Voltammetry (NPV) from the Experiments menu (see Figure 4).
Doing so creates an entry within the archive, called NPV Parameters. In the right pane of the AfterMath application, several tabs will be shown (see Figure 5).
As with most Aftermath methods, the experiment sequence is
Induction Period → Pulse Sequence → Relaxation Period → Post-Experiment Idle Conditions
Continue reading for detailed information about the fields on each unique tab.3.1. Basic Tab
TIP:
Click the AutoFill button on the top bar in AfterMath to automatically fill all required parameters with reasonable starting values. While the values provided may not be appropriate for your specific system, they are reasonable parameters with which to start your experiment, especially if you are new to the method.
The table below lists the group and field names and symbols for each parameter associated with this experiment. See Table 1
Group Name
Field Name
Symbol
Sweep (Sweep limits)
Segments
Sweep (Sweep limits)
Initial Potential
Sweep (Sweep limits)
Initial Direction
Sweep (Sweep limits)
Upper Potential
Sweep (Sweep limits)
Vertex Potential
Sweep (Sweep limits)
Lower Potential
Sweep (Sweep limits)
Final Potential
Normal Pulse (Pulse parameters)
Baseline
Normal Pulse (Pulse parameters)
Width
Normal Pulse (Pulse parameters)
Period
Normal Pulse (Pulse parameters)
Increment
Normal Pulse (Pulse parameters)
Sampling Width
Table 1: Basic Tab Group Names, Field Names, and Symbols.
3.2. Advanced Tab
The NPV Advanced tab contains groups for Induction Period, Relaxation Period, and iR Compensation (see Figure 8).
Induction Period is the first step in an NPV experiment if the Duration is >0 s. During the induction period, the specified current is applied to the cell for the specified duration. During this period, data are not collected. The Induction Period is believed to "calm" the cell prior to intentional perturbation. More on Induction Period is found within the knowledgebase.
Relaxation Period is the last step in an NPV experiment if the Duration is >0 s. During the relaxation period, the specified current is applied to the cell for the specified duration. During this period, data are not collected. The Relaxation Period is believed to "calm" the cell after intentional perturbation. More on Relaxation Period is found within the knowledgebase.
Lastly, the iR Compensation group allows users to adjust the cell feedback to accommodate a known resistive drop between working and reference electrodes. Not all potentiostats from Pine Research support iR compensation. The WaveDriver series and WaveNow Wireless series support iR compensation by positive feedback. The WaveNow, WaveNow XV, WaveNano and the CBP bipotentiostat do not support iR compensation.
The general experimental flow for an NPV experiment is provided below (see Figure 9), highlighting the Induction period, NPV Pulse Sequence, and Relaxation period. Following the relaxation period, the post-experiment conditions are applied.
3.3. Range, Filters, and Post Experiment Conditions Tab
In nearly all cases, the groups of fields on the Ranges tab are already present on the Basic tab. The Ranges tab shows an Electrode Range group and depending on the experiment shows either, or both, current and potential ranges and the ability to select an autorange function. The fields on this tab are linked to the same fields on the Basic tab (for most experiments). Changing the values on either the Ranges tab or on the Basic tab changes the other set. In other words, the values selected for these fields will always be the same on the Ranges tab and on the Basic tab. More on ranges is found within the website.
The Filters tab provides access to potentiostat hardware filters, including stability, excitation, current response, and potential response filters. Pine Research recommends that users contact us for help in making changes to hardware filters. Advanced users may have an easier time changing the automatic settings on this tab.
By default, the potentiostat disconnects from the electrochemical cell at the end of an experiment. There are other options available for what these post-experiment conditions can be and are controlled by setting options on the Post Experiment Conditions tab.
4. Sample Experiment
The typical results for NPV of a 0.5 mM solution of Ferrocene in 0.1 M
show a sigmoid-shaped curve (see Figure 10). In this experiment, the specific parameters were as follows:
show a sigmoid-shaped curve (see Figure 11). In this experiment, the specific parameters were as follows:
- 2 mm OD Pt WE
- Pt mesh CE
- Baseline Potential = 0.1 V
- Initial Potential = 0.25 V
- Final Potential = 1 V
- Pulse increment = 10 mV
- Pulse period = 100 ms
- Pulse width = 10 ms
- Sample width = 1 ms
- 2 mm OD Pt WE
- Pt mesh CE
- Baseline Potential = 1 V
- Initial Potential = 0.75 V
- Final Potential = 0 V
- Pulse increment = 10 mV
- Pulse period = 100 ms
- Pulse width = 10 ms
- Sample width = 1 ms
The typical results for Cyclic Normal Pulse Voltammetry (CNPV) of a 0.5 mM solution of Ferrocene in 0.1 M
show two sigmoid-shaped curves (see Figure 12, same conditions as for those used in generating the data in Figure 11). Since Ferrocene is fully-reversible electrochemically, the two sigmoids overlap.
5. Example Applications
The first example uses NPV to confirm a diffusion coefficient calculated initially by chronoamperometry (CA). Oyaizu et al.3 produced an organic radical polymer to be used in charge-storage applications. The current in this case is controlled by diffusion of the counter ion through the film.
In another example, Welch et al.4 use NPV to measure a diffusion coefficient. They measured the diffusion coefficients of a free and DNA-bound organo-metallic complex. NPV was superior to CV in this instance due to the low concentration of species in solution.
In this third example, Osteryoung et al. use Reverse Normal Pulse Voltammetry (RNPV). The usefulness of RNPV lies in its ability to examine products from chemical reactions that take place after an electrochemical reaction. Osteryoung et al.5 used RNPV to obtain the backwards rate constant and equilibrium constant for the dimerization of N-methyl-2-carbomethoxypyridinium radical, produced after the electrochemical reduction of the N-methyl-2-carbomethoxypyridinium ion.
6. References
- Bard, A. J.; Faulkner, L. A.  Electrochemical Methods: Fundamentals and Applications, 2nd ed. Wiley-Interscience: New York, 2000.
- Kissinger, P.; Heineman, W. R.  Laboratory Techniques in Electroanalytical Chemistry, 2nd ed. Marcel Dekker, Inc: New York, 1996.
- Oyaizu, K.; Ando, Y.; Konishi, H.; Nishide, H.  Nernstian Adsorbate-like Bulk Layer of Organic Radical Polymers for High-Density Charge Storage Purposes.  J. Am. Chem. Soc., 2008, 130(44), 14459-14461.
- Welch, T. W.; Corbett, A. H.; Thorp, H. H.  Electrochemical Determination of Nucleic Acid Diffusion Coefficients through Noncovalent Association of a Redox-Active Probe.  The Journal of Physical Chemistry, 1995, 99(30), 11757-11763.
- Osteryoung, J.; Talmor, D.; Hermolin, J.; Kirowa-Eisner, E.  Reverse pulse voltammetry. Application to second-order following reactions.  The Journal of Physical Chemistry, 1981, 85(3), 285-289.
