Stimulation Options with the WaveNeuro FSCV System

Stimulation Options with the WaveNeuro FSCV System

1. In-Vivo Stimulation

Researchers who use Fast-Scan Cyclic Voltammetry (FSCV) often use classic electrophysiology/neuroscience methods in conjunction with electrochemistry. One such method is current stimulation. By inserting a stimulating electrode and delivering a defined sequence of current pulses, researchers can cause vesicles in synapses to release their contents. Common neurochemicals studied by FSCV include monoamines dopamine, norepinephrine, and serotonin. Measuring changes in the concentration of these neuromolecules, by FSCV, concurrent with electrical stimulation is a common practice in the community. This article describes the options that exist to couple current stimulation with FSCV, using the Pine Research WaveNeuro FSCV Workstations.

2. Current Stimulation Setups

With the WaveNeuro systems, controlled by HDCV Software (distributed, written, and maintained by the University of North Carolina at Chapel Hill) users have two choices for adding current stimulation to their FSCV measurements. Users can either:

  1. Route the stimulus current to the Pine Research headstage cable connector (Passthrough)
  2. Use the stimulus current directly from the isolator

In either case, additional accessories will be required that are not available from Pine Research. There are multiple sources for these accessories. Pine Research does not specifically recommend any one brand over another; however, we have shown the integrations below using common products used by many laboratories.  Both options below assume that you are using a WaveNeuro FSCV Workstation from Pine Research and HDCV SoftwareHDCV Software from UNC.  In HDCV, navigate to the STIM tab under “Set up Experiment” (see Figure below).

HDCV Stim Tab Settings

HDCV Stim Tab Settings

In this screen, users specify the analog stimulus pulse they want to apply per frame. In the plot on the right, the FSCV waveform is overlaid with the analog stimulus pulse. A third curve, called “Error,” is drawn to to indicate if you have any overlap between the pulse sequence and your FSCV waveform, which will cause problems during use. Resolve any red curve overlapping areas by modifying either (or both) the STIM parameters of your CV RAMP parameters.

On the WaveNeuro FSCV Potentiostat, find the area on the top panel labeled STIMULUS (see Figure below). There, you will find two BNC connectors, one for the output signal (STIM OUT, AO3, stimulus sequence as a voltage) and a PASSTHRU BNC, which passively takes an input and routes it to the terminal of the headstage cable stim connector.

WaveNeuro Stimulus Panel Connections

WaveNeuro Stimulus Panel Connections

3. Passthrough Option

One of the advantages of the Pine Research headstage cable kit is that it terminates with two connectors. One connector is used to connect the headstage, flywires, and working and reference electrodes. The other connector is a common two-prong stimulus connector (305/CP from Plastics1/Invivo1/P1 Technologies/Protech International) (see image below). Having the connector physically close to where the working and reference electrodes are available minimizes the challenge of running multiple leads into an already cramped space.

CP305 Two-Prong Stimulation Electrode Connector

CP305 Two-Prong Stimulation Electrode Connector

Current stimulation is accomplished with several integrated aspects:

  1. Users must setup a stimulation pulse sequence in HDCV that agrees with (i.e., does not overlap) the FSCV ramp applied during electrochemical measurements (see above).
  2. The pulse sequence defined in HDCV is output as a voltage, and is mapped to the STIMULUS  STIM OUT (AO3) BNC on the front panel.  This voltage must be delivered to an isolated current stimulus box to be converted to a current.
  3. The current output from the current stimulus box is made available at end headstage termination connector by feeding the signal through the WaveNeuro (i.e., inputting the current to the PASSTHRU (headstage cable) BNC on the front panel.

The components/accessories required to make this setup are shown below (see figure and table below).  There are many sources of these products.  You do not have to use the brands shown in the image.  Each brand may have different types of cables (input and output connectors) that are required, but the list and figure below is the basic setup for feeding the current stimulus through to the headstage cable termination adapter.

Passthrough Stimulation Setup Diagram

Passthrough Stimulation Setup Diagram

Letter Code
Description
Brand
Part/Model # (link)

A
FSCV Potentiostat with Stim Output
Pine Research

B
BNC to Lemo (1) Cable
DigiTimer, Ltd.

C
Current Stimulus Isolator
DigiTimer, Ltd.

D
Two-Lead 1 mm Pin to BNC Adapter
Pomona (DigiKey)

E
BNC (male-to-male) Cable
Many
Widely available

F
Headstage Cable Kit
Pine Research

G
305/CP Extension Cable
Protech International

H
305 Stimulating Electrode
Protech International

WaveNeuro Stimulus Setup with Passthrough

4. Direct Option

The other option is to deliver the stimulus current directly from the current stimulus isolator to the animal. Initial setup is the same as the passthrough option above, but instead of passing the current through the WaveNeuro to the end of the headstage cable, the current goes directly to the animal.  Current stimulation is accomplished with several integrated aspects:

  1. Users must setup a stimulation pulse sequence in HDCV that agrees with (i.e., does not overlap) the FSCV ramp applied during electrochemical measurements (see above).
  2. The pulse sequence defined in HDCV is output as a voltage, and is mapped to the STIMULUS STIM OUT (AO3) BNC on the front panel. This voltage must be delivered to an isolated current stimulus box to be converted to a current.
  3. The current output from the current stimulus box can then be delivered directly to the animal by connecting appropriate cables and stimulating electrode.

The components/accessories required to make this setup are shown below (see figure and table below). There are many sources of these products. You do not have to use the brands shown in the image. Each brand may have different types of cables (input and output connectors) that are required, but the list and figure below is the basic setup for feeding the current stimulus through to the headstage cable termination adapter.

Direct Stimulation Setup Diagram

Direct Stimulation Setup Diagram

Letter Code
Description
Brand
Part/Model # (link)

A
FSCV Potentiostat with Stim Output
Pine Research

B
BNC to Lemo (1) Cable
DigiTimer, Ltd.

C
Current Stimulus Isolator
DigiTimer, Ltd.

D
Phone Tips to 305/CP Cable
Protech International

E
305 Stimulating Electrode
Protech International

WaveNeuro Stimulus Setup Direct to Animal