The WaveNeuro Fast Scan Cyclic Voltammetry (FSCV) Potentiostat System is a unique electrochemical instrument you will not find from other traditional electrochemical suppliers.
[alert color=”green” icon=”exclamation-triangle”]PLEASE NOTE: All specifications are subject to change without notice.[/alert] [toggle title = “Grounding”]Chassis Terminal
The chassis terminal is a banana binding post (back panel) which may optionally be used to connect the chassis to earth ground to improve noise screening (shielding).
Resolution
Aux Input Resolution
Filters
16-bit resolution (provided by the interface board)
16-bit resolution (provided by the interface board)
Selectable: low-pass up to 14 kHz, low-pass up to 40 kHz, bandpass (100 Hz to 14 kHz) or bypass (no filtering)
Resolution
DAC Output
Filters
CV Scan Rate (max)
100 μV (minimum resolution defined by the interface board)
16-bit resolution (provided by the interface board)
Selectable; lowpass up to 2 kHz, 5 kHz or 10 kHz, and bypass (no filtering)
5,000 V/s
Resolution (at each range)
ADC Input
313 μV (minimum resolution defined by the interface board)
16-bit resolution (provided by the interface board)
Resolution (at each range)
ADC Input
313 μV (minimum resolution defined by the interface board)
16-bit resolution (provided by the interface board)
Synchronization
Raw Point Total
simultaneous sampling of all analog input signals
>10 million per experiment
Headstage Cable
Electrode Connections
Power Supply
Headstage Amplifier
DB-25 to Pine Research electrochemical headstage
white = reference; yellow = working
C14-type socket (includes plug-specific 10A cord)
200 nA/V, working electrode driven (other gains available by special order)
All output connections to the WaveNeuro are via female BNC-type connectors.
STIM –
CVF
SPARE
FLOW
E-PHYS
STIM OUT
Digital, TTL compatible
CV frequency; Digital, TTL compatible
Digital, TTL compatible
Flow cell trigger; Digital, TTL compatible
Timing Trigger; Digital, TTL compatible
Stimulus waveform; Analog, max ±10.0 V
BEHAVIORAL
Female DB-25 connector for recording up to 15 TTL signals
Connector 1
Compatible DAQ Interface
DAQ Interface Cable
Control Software
68 –pin VHDCI male connector
National Instruments PCIe-6363 (sold separately)
National Instruments SHC8-68 EPM x 2
Industry standard LabVIEW based software available to researchers (not provided)
Power Cord
LED Indicators
Instrument Dimensions
Instrument Weight
Temperature Range
Humidity Range
Power Supply
various international cables available (10A, C13 Connector)
power, input and output filter selection, trigger activity
58× 242 × 242 mm (2.25 × 9.5 × 9.5 in)
2.0 kg (1.8 lb)
10˚C to 40˚C
80% RH maximum, non-condensing
input: 100 to 240 VAC,2.3 A,50 to 60 Hz
output: 12.0 VDC,2.5 A