      This FAQ applies to versions 1.0.2447 and greater.

Symptoms

Attempting to perform an experiment from AfterMath on the Pine AFCBP1 bipotentiostat using the ISA-type National Instruments card (AT-MIO-16E-10) may result in the following error:

Experiment failed: "<experiment_name_here>". Instrument failure.

More details

This problem can be caused by a lack of DMA resources. This can be confirmed using either AfterMath or the National Instruments Measurement & Automation Explorer as shown below:

atmio16e10-failure-08.jpg

atmio16e10-failure-07.jpg

atmio16e10-failure-09.jpg

The original cause of this failure occurs during the installation of the National Instruments interface board device drivers, and can also be seen in the Windows Device Manager:

atmio16e10-failure-01.jpg

atmio16e10-failure-02.jpg

Solution

The problem can be fixed by using the Windows “Device Manager” as shown below.

See also the National Instruments website article about this problem here

atmio16e10-failure-03.jpg

atmio16e10-failure-04.jpg

atmio16e10-failure-05.jpg

atmio16e10-failure-06.jpg