Square Wave Voltammetry (SWV)

Last Updated: 4/24/20 by Tim Paschkewitz

ARTICLE TAGS
  • AfterMath SWV,
  • osteryoung,
  • square wave,
  • square wave voltammetry,
  • SWV
Article Contents/Section Navigation
  1. Technique Overview
  2. Fundamental Equations
  3. Experimental Setup in AfterMath
    1. Basic Tab
    2. Advanced Tab
    3. Ranges, Filters, and Post Experiment Conditions Tab
  4. Sample Experiments
  5. Example Applications
  6. References

1Technique Overview

Square Wave Voltammetry (SWV) is a potentiostatic method that offers some advantages to common techniques like Cyclic Voltammetry (CV), in that the waveform is a series of pulses increasing along a linear baseline, where current is measured in a forward pulse and reverse pulse.  The result is the current difference between forward and reverse pulses.  The way in which the current is measured at each pulse aids in minimizing the measurement of background (charging) current.  Related pulse-type methods include Normal Pulse Voltammetry (NPV) and Differential Pulse Voltammetry (DPV).

 
SWV combines the aspects of several pulse voltammetric methods, including the background suppression and sensitivity of DPV, Differential Pulse Voltammetry (DPV) the diagnostic value of NPV, Normal Pulse Voltammetry (NPV) and the ability to interrogate products directly in the manner of Reverse NPV. Normal Pulse Voltammetry (NPV)
 
In an SWV experiment, the potential of the working electrode is stepped through a series of forward and reverse pulses from an initial potential to a final potential.  The forward step is determined by the square amplitude and the reverse step is determined by subtracting the square increment from the square amplitude. Cyclic Square Wave Voltammetry (CSWV) is a variant where the potential of the working electrode is cycled between an Upper potential and a Lower potential.
 
In the simplest case, when Segments (SN) = 1 (see Figure 1), potential pulses in a forward and reverse pulse, along a linearly increasing baseline from an initial to final potential, sampling current at specified intervals (see Figure 2).
 
Square Wave Voltammetry (SWV) One Segment Waveform

Figure 1. Square Wave Voltammetry (SWV) One Segment Waveform

 
During each forward or reverse pulse, current is sampled at the same period within the pulse duration as set by entering Sampling width (TSW, W) into the Basic tab (see below). The period of a pulse (TSW, P) includes both forward and reverse pulses; thus, the current sampled in the forward pulse is at,
 
\displaystyle {\frac{1}{2}T_{SW,\; P}-T_{SW, \; W}}
 
and the current is sampled in the reverse pulse at,
 
\displaystyle {T_{SW,\; P}-T_{SW, \; W}}
 
as shown in Figure 2.
 
Square Wave Voltamemtry Forward and Reverse Pulse Sampling Diagram

Figure 2. Square Wave Voltammetry Forward and Reverse Pulse Sampling Diagram

 
In a later Section (Experimental Setup in AfterMath - Basic Tab), the pulse parameters are defined in detail (see Table 1 and Figure 7).
 
When Segments (SN) = 2 (see Figure 3), potential pulses in a forward and reverse pulse, along a linearly increasing baseline from an initial potential to vertex potential and then to final potential, sampling current at specified intervals (see Figure 2).
 
Square Wave Voltammetry (SWV) Two Segment Waveform

Figure 3. Square Wave Voltammetry (SWV) Two Segment Waveform

 
When Segments (SN) = 3 (see Figure 4), potential pulses in a forward and reverse pulse, along a linearly increasing baseline from an initial potential to upper potential to lower potential and then to final potential, sampling current at specified intervals (see Figure 2).
 
Square Wave Voltammetry (SWV) Three Segment Waveform

Figure 4. Square Wave Voltammetry (SWV) Three Segment Waveform

Back To Top

2Fundamental Equations

The following is a brief introduction to the theory of SWV.  SWV was invented by Ramaley and Krause Krause, M. S. ; Ramaley, L.   Analytical application of square wave voltammetry.  Anal. Chem., 1969, 41(11), 1365-1369. and thoroughly explained and further developed by the Osteryoungs. Osteryoung, J. G. ; Osteryoung, R. A.   Square wave voltammetry.  Anal. Chem., 1985, 57(1), 101-110. Refer to these references, including Bard and Faulkner's Electrochemical Methods for additional details on Square Wave Voltammetry. Bard, A. J. ; Faulkner, L. R.   Electrochemical Methods Fundamentals and Applications, 2nd ed. John Wiley & Sons, Inc.: New York, 2001. Cyclic Square Wave Voltammetry (CSWV), originally developed by Xinsheng and Guogang, Xinsheng, C. ; Guogang, P.   Cyclic Square Wave Voltammetry: Theory and Experimental.  Anal. Lett., 1987, 20(10), 1511-1519. and then revived by Helfrick and Bottomley, Helfrick, J. C. ; Bottomley, L. A.   Cyclic Square Wave Voltammetry of Single and Consecutive Reversible Electron Transfer Reactions.  Anal. Chem., 2009, 81(21), 9041-9047. is covered in the literature also.
 
Consider a reaction O + e^- \rightarrow R, where  O is reduced in a one-electron reaction to R with formal potential, E^{0}.  An initial potential is applied to the electrode that is significantly more positive than E^{0}.  No significant faradaic current flows upon the application of forward pulse towards more positive values.  Current is sampled at a specified point during the forward pulse.
 
The reverse pulse consists of stepping the potential of the working electrode to more negative values and the current is sampled near the end of the reverse pulse.  The difference current is calculated by subtracting the reverse current from the forward current, given as
 
i_d=i_{forward}-i_{reverse}
 
As the potential of the working electrode approaches E^{0} a faradaic current flows due to reduction of  O .  Upon application of the reverse pulse, a faradaic current flows that effectively oxidizes the Rthat was produced during the forward pulse.  In other words, the rate of reduction slows compared to the forward step and an anodic current flows.  Once the potential of the working electrode is sufficiently more negative of E^{0} the current is diffusion-limited in both the forward and reverse pulse directions and the difference current is small.
 
The strength of SWV lies in its diagnostics, meaning that it is not typically used for quantitative analysis.  It is possible; however, to calculate peak height using the equation
 
\displaystyle {{\Delta}i_p = \frac{nFAD_O^{1/2}C_O^*}{{\pi}^{1/2}t_p^{1/2}}{\Delta}{\psi}}
 
where n is the number of electrons, F is Faraday's Constant (96,485 C/mol), A is the electrode area (cm2), D_{0} is the diffusion coefficient of species O (cm2/s), C_O is the concentration of species O (mol/cm3), t_p is the experimental time scale \displaystyle {1/2T_{SW,\; P}-T_{SW, \; W}}, and {\Delta}{\psi}_p is a dimensionless peak Current parameter (see Table 1).
 
n \Delta E_{A,\; SWV}/mV n \Delta E_{I,\; SWV}/mV
1 5 10 20
0* 0.0053 0.0238 0.0437 0.0774
10 0.2376 0.2549 0.2726 0.2998
20 0.4531 0.4686 0.4845 0.5077
50 0.9098 0.9186 0.9281 0.9432
100 1.1619 1.1643 1.1675 1.1745
Table 1.  Dimensionless Peak Current ({\psi}_p) vs. SWV Operating Parameters
 
The data in Table 1 were obtained in Electrochemical Methods. Bard, A. J. ; Faulkner, L. R.   Electrochemical Methods Fundamentals and Applications, 2nd ed. John Wiley & Sons, Inc.: New York, 2001. .  *In Table 1, n \Delta E_{I,\; SWV}/mV = 0 corresponds to staircase voltammetry. Staircase Voltammetry (SCV)
Back To Top

3Experimental Setup in AfterMath

To perform a square wave voltammetry experiment in AfterMath, choose Square Wave Voltammetry (SWV) from the Experiments menu (see Figure 5).
 
Square Wave Voltammetry (SWV) Experiment Menu Selection in AfterMath

Figure 5. Square Wave Voltammetry (SWV) Experiment Menu Selection in AfterMath

 
Doing so creates an entry within the archive, called SWV Parameters. In the right pane of the AfterMath application, several tabs will be shown (see Figure 6).
 
Square Wave Voltammetry (SWV) Experiment Basic Tab

Figure 6. Square Wave Voltammetry (SWV) Experiment Basic Tab

 
As with many Aftermath methods, the experiment sequence is
 
Induction Period → Square Wave Sequence → Relaxation Period → Post-Experiment Idle Conditions
 
Continue reading for detailed information about the fields on each unique tab.
Back To Top

3.1Basic Tab

TIP:  Click the AutoFill button ("I Feel Lucky" prior to May 2019) on the top bar in AfterMath to automatically fill all required parameters with reasonable starting values.  While the values provided may not be appropriate for your specific system, they are reasonable parameters with which to start your experiment, especially if you are new to the method.

The basic tab contains fields for the fundamental parameters necessary to perform a SWV experiment.  AfterMath shades fields with yellow when a required entry is blank and shades fields pink when the entry is invalid (see Figure 7).
 
Square Wave Voltammetry (SWV) Basic tab in AfterMath

Figure 7. Square Wave Voltammetry (SWV) Basic tab in AfterMath

 
During the induction period, Induction Period a set of initial conditions are applied to the electrochemical cell and the cell equilibrates at these conditions.  Data are not collected during the induction period, nor are they shown on the plot during this period.  Users will define induction period parameters on the Advanced Tab.
 
After the induction period, the potential of the working electrode is stepped through a series of increasing forward and reverse pulses from the Initial potential to the Final potential.  The potential is incremented with each successive pulse according to the Pulse increment.  In a typical experiment, Segments = 1 and there is a single series of pulses moving along the linear baseline from initial to final potential.  Some may want to reverse the pulses (move in the opposite direction), accomplished by adjusting the number of segments to be > 1.  By adjusting the number of segments, users can create variants of the DPV technique.  Cyclic Square Wave Voltammetry consists of cycling (through a series of potential pulses also) the potential of the working electrode through a Vertex potential or between Upper potential and a Lower potential until reaching Final potential (Segments = 2 or 3).
 
The parameter settings for pulse-type experiments Pulse Methods are often not as clear as with something more simple like Cyclic Voltammetry (a sweep method ).  Refer to Figure 8 below when understanding each parameter of the DPV pulse.  Further, clicking "AutoFill" ("I Feel Lucky" in older versions of AfterMath) provides reasonable starting parameters if you are unsure of typical values used in these experiments.  Most often, research journal articles describe the SWV pulse sequence used, which can be replicated using AfterMath.
 
The experiment concludes with a relaxation period.  Relaxation Period  During the relaxation period, a set of final conditions (specified on the Advanced tab) are applied to the electrochemical cell and the cell equilibrates at these conditions (set on the Advanced Tab).  Data are not collected during the induction period, nor are they shown on the plot during this period.
 
At the end of the relaxation period, the post-experiment idle conditions are applied to the cell and the instrument returns to the idle state.
 
A plot of the typical experiment sequence, containing labels of the fields on the Basic tab, helps to illustrate the sequence of events in an SWV experiment (see Table 1 and Figure 8).
 

The table below lists the group and field names and symbols for each parameter associated with this experiment (see Table 1).

Group Name Field Name Symbol
Sweep (Sweep limits) Segments S_N
Sweep (Sweep limits) Initial Potential E_I
Sweep (Sweep limits) Initial Direction DIR_I
Sweep (Sweep limits) Upper Potential E_U
Sweep (Sweep limits) Vertex Potential E_V
Sweep (Sweep limits) Lower Potential E_{LOW}
Sweep (Sweep limits) Final Potential E_F
Square Wave Amplitude E_{A, \;SWV}
Square Wave Period E_{P, \;SWV}
Square Wave Increment E_{E, \;SWV}
Square Wave Sampling width E_{W, \;SWV}
Table 1.  Basic Tab Group Names, Field Names, and Symbols.
 
Square Wave Voltammetry (SWV) Pulse Sequence Detail

Figure 8. Square Wave Voltammetry (SWV) Pulse Sequence Detail

Back To Top

3.2Advanced Tab

The SWV Advanced tab contains groups for Induction Period, Relaxation Period, Pulse options, and iR Compensation (see Figure 9).
 
Square Wave Voltammetry (SWV) Experiment Advanced Tab in AfterMath

Figure 9. Square Wave Voltammetry (SWV) Experiment Advanced Tab in AfterMath

 
Induction Period is the first step in a SWV experiment if the Duration is >0 s.  During the induction period, the specified current is applied to the cell for the specified duration.  During this period, data are not collected.  The Induction Period is believed to "calm" the cell prior to intentional perturbation (see Figure 10).  More on Induction Period is found within the knowledgebase. Induction Period
 
Relaxation Period is the last step in a SWV experiment if the Duration is >0 s.  During the relaxation period, the specified current is applied to the cell for the specified duration.  During this period, data are not collected.  The Relaxation Period is believed to "calm" the cell after intentional perturbation (see Figure 10).  More on Relaxation Period is found within the knowledgebase. Relaxation Period
 
Square Wave Voltamemtry (SWV) Advanced Tab Fields

Figure 10. Square Wave Voltamemtry (SWV) Advanced Tab Fields

 
Detailed description of the iR Compensation Mode is provided elsewhere on the knowledgebase.  iR Compensation This mode is used to correct for uncompensated resistance in the electrochemical cell.
 
Lastly, the iR Compensation group allows users to adjust the cell feedback to accommodate a known resistive drop between working and reference electrodes.  Not all potentiostats from Pine Research support iR compensation.  The WaveDriver series support iR compensation by positive feedback and current interrupt.  The WaveDriver 100 WaveDriver 100 EIS Potentiostat Basic Bundle and WaveDriver 200 WaveDriver 200 EIS Bipotentiostat Basic Bundle support EIS-based iR compensation.  The WaveNow WaveNow Low-Voltage Bundle series (including the WaveNano WaveNano Low-Current Bundle and WaveNowXV WaveNowXV Potentiostat/Galvanostat Basic Bundle ) and the CBP bipotentiostat do not support iR compensation of any type.  More information about iR compensation, including understanding how it works and how to determine the resistance, consult the knowledgebase article on the topic. iR Compensation
Back To Top

3.3Ranges, Filters, and Post Experiment Conditions Tab

In nearly all cases, the groups of fields on the Ranges tab are already present on the Basic tab.  The Ranges tab shows an Electrode Range group and depending on the experiment shows either, or both, current and potential ranges and the ability to select an autorange function.  The fields on this tab are linked to the same fields on the Basic tab (for most experiments).  Changing the values on either the Ranges tab or on the Basic tab changes the other set.  In other words, the values selected for these fields will always be the same on the Ranges tab and on the Basic tab.  More on ranges is found within the knowledgebase, Electrode Range as is for autorange. Autorange
 
The Filters tab provides access to potentiostat hardware filters, including stability, excitation, current response, and potential response filters.  Pine Research recommends that users contact us Contact  for help in making changes to hardware filters.  Advanced users may have an easier time changing the automatic settings on this tab.  Filter settings fields are shown for WK1 (working electrode #1) as well as for WK2 (working electrode #2) regardless of the potentiostat connected to AfterMath.  More information on filters is available elsewhere on the knowledgebase. Hardware Filters
 
By default, the potentiostat disconnects from the electrochemical cell at the end of an experiment.  There are other options available for what these post-experiment conditions can be and are controlled by setting options on the Post Experiment Conditions tab.  Complete details on the fields and settings on this Post Experiment Conditions tab are provided elsewhere on the knowledgebase. Understanding Post Experiment Conditions Tab
Back To Top

4Sample Experiments

Below are the typical results for SWV (current difference) for the oxidation of a 1 mM solution of ferrocene n 0.1 M Bu4NClO4/CH2CL2 (see Figure 11).  The specific parameters for this experiment are as follows:
 
  • Amplitude: 25 mV
  • Period: 10 ms
  • Increment: 2 mV
  • Sampling width: 1 ms
 
Square Wave Voltammogram (Difference Current) of a Ferrocene Solution

Figure 11. Square Wave Voltammogram (Difference Current) of a Ferrocene Solution

 
As mentioned previously, the default output plot from a SWV experiment is current difference (between forward and reverse pulses) vs. applied potential, as shown (see Figure 11). Under "Other Plots" there are square wave voltammograms for the forward pulse (see Figure 12) and reverse pulse (see Figure 13).
 
Square Wave Voltammogram (Forward Pulse Only) of a Ferrocene Solution

Figure 12. Square Wave Voltammogram (Forward Pulse Only) of a Ferrocene Solution

 
Square Wave Voltammogram (Reverse Pulse Only) of a Ferrocene Solution

Figure 13. Square Wave Voltammogram (Reverse Pulse Only) of a Ferrocene Solution

 
Below is an example of Cyclic Square Wave Voltammetry, when Segments (SN) > 1 (see Figure 14).  Peak potentials are marked with a Crosshair tool to show that the anodic and cathodic peaks appear at nearly the same potential.
 
Cyclic Square Wave Voltammogram (Difference Current) for a Ferrocene Solution

Figure 14. Cyclic Square Wave Voltammogram (Difference Current) for a Ferrocene Solution

Back To Top

5Example Applications

The first application uses SWV to monitor binding events associated with electrochemical sensors.  White and Plaxco White, R. J.; Plaxco, K. W.  Exploiting Binding-Induced Changes in Probe Flexibility for the Optimization of Electrochemical Biosensors.  Anal. Chem., 2010, 82(1), 73-76. developed redox-tagged electrochemical sensors from electrode-bound oligonucleotides. Tuning the frequency of the voltammetric measurements allows the researchers to amplify both the unbound and target-bound signals, essentially, making an “On/Off” sensor.
 
In another example, Chakrabarti et. al Chakrabati, C.; Cheng, J.; Lee, W. F.; Back, M.; Schroeder, W.  Rotating Disk Electrode Voltammetry for Studying Kinetics of Metal Complex Dissociation in Model Solutions and Snow Samples.  Environ. Sci. Technol., 1996, 30(4), 1245-1252. used several electrochemical techniques, SWV among them, to monitor metal dissociation events.  They combined RDE Rotating Disk and Rotating Cylinder Electrode Setups (RDE and RCE)
Rotating Methods
Hydrodynamic Electrochemistry
Rotating Disk Voltammetry (RDE)
Rotating Ring-Disk Voltammetry (RRDE)
with anodic stripping voltammetry to distinguish between labile and nonlabile complexes in extremely low concentrations in aqueous solutions and in precipitation samples.  The authors then showed that DPV, Differential Pulse Voltammetry (DPV) SCV, Staircase Voltammetry (SCV) and SWV each provide similar dissociation constants; however, the sensitivity of SWV was two orders of magnitude higher than SCV.  This is a nice example of how SWV compares with other electrochemical techniques.
Back To Top

6References

Back To Top
Questions? Find Answers.

Our knowledgebase is the central repository for written content, including help topics, theory, application notes, specifications, and software information.

Select a category below or view the Knowledgebase home page
Software

Detailed information about our Software, which includes AfterMath and retired PineChem.

Applications

Application notes discuss practical aspects of conducting specific experiments.

Theory

Fundamental electrochemical theory presented in a brief and targeted manner.

Product Specifications

Review complete product specifications and compare products within a category here.